Safety operation rules for rotary vane vacuum pump
I. purpose and scope
of application of rotary vane vacuum pump
1. Vacuum pump is a basic
equipment to pump out the gas in a specific sealed container and make the
container obtain a certain degree of vacuum.
2. The vacuum pump is not
suitable for pumping all kinds of gases with high oxygen content, toxic,
explosive etched ferrous metal and chemical effect on vacuum oil, nor can it be
used as compressor and delivery pump.
3. Vacuum pump is generally used
at room temperature of 5-40 ℃ and relative humidity not higher than 90%.
II. Operating instructions of rotary vane vacuum pump
1. Check
whether the power supply is connected, and start the vacuum pump motor after
confirming the connection is correct.
2. Start 1-2 times intermittently,
observe whether there is abnormal sound and special vibration during operation,
and operate continuously only when there is no problem.
3. Check whether
the cooling water pipe is smooth.
4. The air inlet of the pump shall be
operated continuously in the open air for no more than three minutes to prevent
the site from being polluted by fuel injection.
5. Pay attention to the
position of the oil surface. Too much oil will cause poor start-up, oil return,
oil injection and other adverse phenomena. Too little oil will not play the role
of oil seal on the exhaust valve. There will be a large exhaust sound which will
affect the vacuum degree. When the oil is insufficient, oil should be added
through the oil filling hole, and the center of the oil injection pump oil
standard is appropriate when the pump is stopped.
6. Vacuum pump oil of
different types and brands shall not be mixed.
7. Open the valve on the
intake nozzle of the pump, and the opening size depends on the specific
situation.
8. When the pump is stopped, the valve on the air inlet
nozzle shall be closed first. If the air vent valve is installed, the pump body
shall be vented, and then the power supply shall be cut off.
9.
Precautions
(1) always pay attention to the oil level;
(2)
slowly open the valve after the pump is started;
(3) always pay
attention to whether the pump operates normally, whether there is special sound,
and whether the motor operates under overload;
(4) the cooling water
shall not be interrupted;
(5) pay attention that the temperature rise of
the pump does not exceed 40 ℃.